二氧化氯ClO2气体检测管116 CHLORINE DIOXIDE 1-20ppm
Tube No.
116 CHLORINE DIOXIDE
1. PERFORMANCE
1) Measuring range :1-20 ppm
Number of pump strokes 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :0.3 ppm
4) Shelf life :2 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Concentration chart method
8) Colour change :White→Reddish orange
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
0-Toluidine is oxidized and Orthoquinone is produced.
4. CALIBRATION OF THE TUBE
COLOURIMETRY METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
Substance Interference ppm Coexistence
Nitrogen dioxide FIG.1 Similar stain is produced. 1 Higher readings are given.
Chlorine FIG.2 〃1 〃
Bromine FIG.3 〃1 〃