Technical specifications R.I.T.A.
Linearity of transducer (r2.5mm): r0.1% Full scale
Reproducibility: r0.5%
Accuracy: r2%, typical r1% at samplelength
of 10..20mm
Zero adjustment (by motor): 0..30mm
Range: r50...2500Pm
Output voltage of amplifier : r1.0V Full scale
Temperature drift of amplifier: r0.1Pm/K
Temperature drift of transducermounting: r0.5Pm/K
Sample holder: Fused silica
Sample size: Inner diameter: 5mm
(hollow sample) Outer diameter: 6...7mm
Length: 10mm
Temperature range: RT...1000°C
Thermocouple: Type-K (NiCr/Ni), welded to
sample
Heating: Inductive
Frequency range Generator: 150..400kHz
Heating/Cooling rate: 0.1..99.9K/s
0.1..99.9K/min
Dwell time: 0...3000s
0..3000min
Acquisition rate: max. 1000 readings/s
Power requirements: 230VAC, 16A, 50..60Hz
Power consumption Generator: 3.5kVA
Cooling water: 2l/min, 5 Bar
Sample/Cooling gas: Inert gas, 10l/min max. at 1 Bar max.